Item shown is representative of range - Catalogue No.:NG45-EXC001

Item shown is representative of range - Catalogue No.:NG45-EXC001

Excel Healthcare - ENIGMA NEW GENERATION 45LB

Stock Code: NG45-EXC001

This part number has been discontinued, please contact our sales team for the closest alternative.

This product has been discontinued by the manufacturer. Please contact us for the closest replacement - Autoclave Excel Healthcare ENIGMA NEW GENERATION 45LB benchtop, 45l capacity, suitable for liquids, user programmable, front loading, no autofill,...

read more

List Price: Offer Price: Quantity:
£0.00 £6,586.00
List Price: £0.00
Offer Price: £6,586.00
Quantity:
Manufactured in uk1 Year WarrantyDelivery Time

This part number has been discontinued, please contact our sales team for the closest alternative.

This product has been discontinued by the manufacturer. Please contact us for the closest replacement - Autoclave Excel Healthcare ENIGMA NEW GENERATION 45LB benchtop, 45l capacity, suitable for liquids, user programmable, front loading, no autofill, vacuum air removal, 1 chamber, usable depth 620mm, usable diameter 319mm, electric steam generator heating method, with temperature and pressure gauges, printer and USB included, suitable for laboratory, medical and dental use


Power Draw:3.00kW
Type:Benchtop
Capacity:45l
Suitable for liquids?:Yes
User programmable:Yes
Loading:Front
Autofill:No
Vacuum air removal:Yes
Number of chambers:1
Heat source required:No
Chamber shape:Round
Usable width:0mm
Usable height:0mm
Usable depth:620mm
Usable diameter:319mm
Heating method:Electric steam generator
Max run temperature:134°C
Automatic door opening:Yes
Sliding door:No
Temperature and pressure gauges included:Yes
Data output:Printer and USB
Disinfection type:Steam
Assisted air cooling:No
Suitable for laboratory use:Yes
Suitable for dental use:Yes
Suitable for medical use:Yes
Suitable for beauty/tattoo use:No

Minimum specification that this product meets

  • The chamber to be constructed from high-grade stainless steel conforming to PD5500:2009 as amended
  • The general design, construction, operation and performance should conform to BS2646-1:1993 and/or BS EN 13060:2004 as amended.
  • Capable of generating its own steam electrically. Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363
  • Chamber access via a door fitted with a safety mechanism that allows any residual pressure in the chamber to vent safely before the door can be opened fully.
  • Control panel, incorporating a chamber pressure gauge, allowing the user to select appropriate sterilisation cycles and monitor the progress of the sterilisation cycle via a digital/graphical display.
  • A minimum of 4 user selectable cycle programmes held in the autoclaves memory
  • The facility to lock cycle settings to prevent inadvertent changes to the settings by infrequent users. A key or PIN accessed supervisor mode would be necessary to set the parameters for the cycles.
  • User adjustable sterilisation temperatures in the range 100oC to 137oC. User adjustable sterilisation times, activated by the achievement of the preset sterilisation temperature, in the range of 1
  • 30 minutes.
  • User adjustable free-steaming time in the range of 0
  • 30 minutes
  • A thermal lock, with key override, on the chamber door that prevents opening before the chamber and goods cool to 80oC