Item shown is representative of range - Catalogue No.:30130868

Item shown is representative of range - Catalogue No.:30130868

Ohaus - Frontier FC5515R Micro

Stock Code: 30130868

Microfuge Ohaus Frontier FC5515R Micro, no rotor included, minumum tube volume 1.5ml, maximum tube volume 5ml, 12 maximum tube capacity, maximum g force 21953, minimum temperature -20°C, maximum temperature 40°C, with run counter

read more

List Price: Offer Price: Quantity:
£4,490.00 £3,791.00
List Price: £4,490.00
Offer Price: £3,791.00
Quantity:
Manufactured in european2 Years WarrantyDelivery Time

Microfuge Ohaus Frontier FC5515R Micro, no rotor included, minumum tube volume 1.5ml, maximum tube volume 5ml, 12 maximum tube capacity, maximum g force 21953, minimum temperature -20°C, maximum temperature 40°C, with run counter


Power Draw:0.50kW
Rotor included:No
Min tube volume:1.5ml
Max tube volume:5ml
Max tube capacity:12
Max g force:21953g
Min temp:-20°C
Max temp:40°C
Run counter:Yes
Temp display:Yes
Rotors suitable for pre cooling available:Yes
Bioseal rotor available as an accessory:Yes

Minimum specification that this product meets

  • A lid interlock system that prevents user access to the rotor whilst it is in motion and prevents the rotor turning whilst the lid is open
  • The rotor bowl, lid and external casing must be sufficiently robust to be able to contain the rotor and all samples therein in the event a rotor or drive spindle failure
  • A variable speed controller with a controllable rotor speed range of 1,000rpm to 12,000rpm as a minimum
  • An adjustable timer to control the duration of centrifugation with a minimum range of 0 to 15 minutes
  • A cooling system that does not utilise CFC refrigerants
  • Adjustable rotor and rotor bowl temperature with a minimum range of +2oC to +40oC
  • A hold feature to allow continuous operation of the machine independent of the timer
  • Straightforward, self-evident controls requiring the minimum of training
  • Legible digital display(s) capable of showing set time, temperature and speed and actual time, temperature and speed
  • Interchangeable rotors to accept, as a minimum, 12 x 1.5ml microcentrifuge tubes or 12 x 0.75ml microcentrifuge tubes
  • Rotor imbalance detection and warning
  • Over speed detection and warning
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363