Item shown is representative of range - Catalogue No.:BM11200E

Item shown is representative of range - Catalogue No.:BM11200E

Contained Air Solutions (CAS) - BioMAT1 1200E

Stock Code: BM11200E

Wolf special offer - Class I microbiological safety cabinet Contained Air Solutions (CAS) BioMAT1 1200E, 1200mm wide for ducted applications, exhaust air flow rate 0.13m3, noise level 54dB, compliant to EN12468:2000 and BS5726:1992 OFFER ENDS 30-04-2...

read more

List Price: Offer Price: Quantity:
£10,793.00 £8,667.00
List Price: £10,793.00
Offer Price: £8,667.00
Quantity:
Manufactured in uk3 Years WarrantyDelivery Time

Wolf special offer - Class I microbiological safety cabinet Contained Air Solutions (CAS) BioMAT1 1200E, 1200mm wide for ducted applications, exhaust air flow rate 0.13m3, noise level 54dB, compliant to EN12468:2000 and BS5726:1992 OFFER ENDS 30-04-2024


Power Draw:0.18kW
Width:1200mm
Type:Ducted
Exhaust airflow rate:0.13m³/s
Noise level:54dB

Minimum specification that this product meets

  • Compliance with the requirements of BS5726:2005/BS EN 12469:2000 and other current, relevant standards
  • Stainless steel (minimum grade AISI 316) or epoxy coated work surfaces
  • MSC to have conspicuous visual indication that the unit is operating safely and an audible & visual alarm indicating any condition that may render the MSC unsafe to use.
  • Air flow rates of at least 0.7 metres/second but no greater than 1.0 metres/second (average across the face) inflow.
  • A variety of widths
  • An inlet to allow for oil aerosol penetration tests of the HEPA filters and seals
  • An additional HEPA exhaust filter (non-ducted models only)
  • Internal lighting in the workspace of at least 1,000 Lux
  • A site visit by a trained operative to advise on the suitability of the proposed location of the MSC.
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363

Desirable features of this product

  • A night door, where necessary
  • A bespoke frame, with adjustable feet, for floor mounting options
  • An option to have short-wave UV lamps fitted (N.B. if this option is chosen, no acrylic materials nor any other materials adversely affect by UV radiation should be fitted within the working area)
  • Internal natural gas valve for connection to an external source
  • Internal vacuum valve for connection to an external source
  • Internal 13A electrical socket outlets (robust and moisture resistant)

Desirable features this product does not have

  • Motorised sash
  • Fumigation kit utilising either formaldehyde or Hydrogen peroxide (with a neutralisation capability for non-ducted MSC)
  • Integral footrest