Item shown is representative of range - Catalogue No.:BM2S21800R

Item shown is representative of range - Catalogue No.:BM2S21800R

Contained Air Solutions (CAS) - BioMAT2 S21800R

Stock Code: BM2S21800R

Class II microbiological safety cabinet Contained Air Solutions (CAS) BioMAT2 S21800R working width 1620mm, with motorised sliding sash, dual fan, for recirculating applications. This unit is preconfigured with two HEPA filters in the exhaust compart...

read more

List Price: Offer Price: Quantity:
£16,410.00 £14,168.00
List Price: £16,410.00
Offer Price: £14,168.00
Quantity:
Manufactured in uk3 Years WarrantyDelivery Time

Class II microbiological safety cabinet Contained Air Solutions (CAS) BioMAT2 S21800R working width 1620mm, with motorised sliding sash, dual fan, for recirculating applications. This unit is preconfigured with two HEPA filters in the exhaust compartment


Power Draw:0.35kW
Ecological Features:Certified carbon neutral.
Working width:1690mm
Front closure:Motorised sliding sash
Number of fans:2
Max power draw:0.57kW
Noise level:60dB
VHP ports?:Yes
Ducted or recirculating:Recirculating
Angled HEPA filter?:Yes
Side Walls:Glass
Number of access ports:4
Warranty:3Years
Working height:680mm
Working depth:395mm

Minimum specification that this product meets

  • Compliance with the requirements of BS5726:2005/EN 12469:2000 and other current, relevant standards
  • Stainless steel (minimum grade AISI 316) or epoxy coated work surfaces
  • MSC to have conspicuous visual indication that the unit is operating safely and an audible & visual alarm indicating any condition that may render the MSC unsafe to use.
  • Air flow rates of at least 0.4 metres/second inflow (average across the face) and a down flow over the working area of at least 0.25 metres/second but no greater than 0.5 metres/second.
  • A variety of widths
  • An inlet to allow for oil aerosol penetration tests of the HEPA filters and seals
  • An additional HEPA exhaust filter (non-ducted models only)
  • Internal lighting in the workspace of at least 1,000 Lux
  • A site visit by a trained operative to advise on the suitability of the proposed location of the MSC.
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363