LTE Scientific -

Stock Code: TCR/40/H1

40 Front Loading Autoclave, In-Chamber Heaters

read more

List Price: Offer Price: Quantity:
£13,545.00 £12,020.00
List Price: £13,545.00
Offer Price: £12,020.00
Quantity:
Delivery Time

40 Front Loading Autoclave, In-Chamber Heaters


Minimum specification that this product meets

  • The chamber to be constructed from high-grade stainless steel conforming to PD5500:2009 as amended
  • The general design, construction, operation and performance should conform to BS2646-1:1993 and/or BS EN 13060:2004 as amended.
  • Capable of generating its own steam electrically. Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363
  • Chamber access via a door fitted with a safety mechanism that allows any residual pressure in the chamber to vent safely before the door can be opened fully.
  • Control panel, incorporating a chamber pressure gauge, allowing the user to select appropriate sterilisation cycles and monitor the progress of the sterilisation cycle via a digital/graphical display.
  • A minimum of 4 user selectable cycle programmes held in the autoclaves memory
  • The facility to lock cycle settings to prevent inadvertent changes to the settings by infrequent users. A key or PIN accessed supervisor mode would be necessary to set the parameters for the cycles.
  • User adjustable sterilisation temperatures in the range 100oC to 137oC. User adjustable sterilisation times, activated by the achievement of the preset sterilisation temperature, in the range of 1
  • 30 minutes.
  • User adjustable free-steaming time in the range of 0
  • 30 minutes
  • A thermal lock, with key override, on the chamber door that prevents opening before the chamber and goods cool to 80oC