Shaking incubator IKA KS 4000 IC Control benchtop, not refrigerated, max vessel capacity 1000ml, top loading, 20mm orbit, not stackable, platform required and not included, max shaking capacity 90,000ml


Power Draw:0.56kW
Benchtop or floor standing:Benchtop
Refrigerated:Optional
Max vessel capacity:1000ml
Platform access:Top loading
Max orbit:20mm
Programmable:Yes
Stackable?:No
Block or platform included?:No
Type of shaking:Orbital
Max shaking capacity:90000ml
External width:580mm
External depth:750mm
Minimum temperature:26°C
Maximum temperature:80°C`
Maximum RPM:500RPM
With antimicrobial protection:No

Minimum specification that this product meets

  • Operate at above ambient, from 5oC above ambient to 70oC
  • A maximum variance across all areas of the incubator chamber of ?0.5oC at any set-point of 20oC (type 1) and 37oC (type 2)
  • An accuracy of ?0.1oC of the displayed temperature
  • An adjustable orbital platform speed control between 25rpm and 400rpm with a deviation of speed of ?1% of the set point
  • An easily accessible and simple to use controller with digital set points and temperature and orbital speed displays.
  • An audible and visible, adjustable high and low temperature alarm
  • A range of interchangeable orbital platform with clamping devices to accommodate range of sizes of vessels
  • Door opening mechanisms that require the minimum of effort and incorporates a safety interlock that stops the orbiting platform when open
  • Overload sensors on the orbital platform to make user aware of overload condition and restrict orbiting speed to a safe level
  • All motor and platform bearings are to be sealed to prevent ingress of liquid in the event of a spillage
  • CFC free refrigerants on sub-ambient models
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363